During first-time use and/or after hard resetting the inhaler, you may perform a short inhalation training session to help you learn the optimal method for inhaling from the SyqeAir Inhaler.
Attention: the training inhalations are for training purposes only, and no active ingredient is transferred to you during these inhalations.
Attention: make sure the battery is charged before using the inhaler. If the battery is not charged, charge it according to the instructions.

Following language selection, the Training Mode Selection screen appears; at this point, you should select ‘Yes’ if you wish to perform a training inhalation session.
For confirmation and operation of training mode, press the plus (+) button. To continue without the training mode, press the minus (-) button. After confirmation of the training mode, the ‘Training Mode activated’ screen will appear. Throughout the training mode, the front indication light will turn purple, and an asterisk symbol (i.e., *) will appear in the upper left corner of the screen.

The ‘Ready to INHALE’ [training] screen will now appear. Training mode consists of 4 training inhalations and will display in which of the 4 inhalations you are currently at.
To exit the training mode at any stage, hold the ON button pressed for 4 seconds until the inhaler shuts off. To activate the inhaler again, repeat the instructions in sections 10.1-10.3.
Exhale before performing an inhalation.
After exhaling, place your lips on the mouthpiece and inhale in a uniform manner for approximately 3 seconds. Towards the end of the inhalation, you will feel resistance in the airflow, but do not stop inhaling until the inhalation indicator light turns off.

The inhaler indicator light will turn off at the end of a successful inhalation, and the ‘SUCCESSFUL Inhalation’ [training] screen will appear.