SyqeAir Inhaler Support

Inhaling Correctly

Attention: Inhale while sitting (or at least in a raised position of 45. ̊
Do not block the air intake.
Do not rotate or remove the mouthpiece during inhalation.

This action may cause damage to the inhaler.

 

 

After exhaling, place your lips on the mouthpiece and inhale at a uniform rate for at least 3 seconds. Towards the end of the inhalation, you will feel a resistance in the airflow, but do not stop inhaling until the inhalation indicator light turns off.

At the end of a successful inhalation, the inhalation indicator light will turn off, and the ‘Successful Inhalation’ screen will appear.

 

 

In the event of an unsuccessful inhalation, 3 beeps will be heard, the inhalator indicator light will turn Orange, and one of the following screens will appear:

 

   A. ‘Weak Inhalation’ screen

Meaning – Attention, Inhalation is incorrect; the desired dosage was not delivered.

 

 

   B. ‘Short Inhalation’ screen

Meaning – Attention, the inhalation period was shorter than 3 seconds, a partial dosage was delivered, and therefore you should wait 20 minutes and assess whether there is a need for an additional inhalation.

 

  C. ‘Forceful Inhalation’ screen

Meaning – Attention, Inhalation was too forceful, and the desired dosage was not delivered.

 

  D. ‘Unsteady Inhalation’ screen

Meaning – Attention, the inhalation was not uniform, and the desired dosage was not delivered.

 

If an incorrect inhalation occurs, one of the two following instruction screens will be displayed:

 

  A. ‘Wait 20 minutes for an additional inhalation’ screen

A ‘Wait 20 minutes’ screen is displayed when an incorrect inhalation has been detected and a partial dose has been delivered. Wait 20 minutes and assess whether there is a need for an additional inhalation.

 

   B. ‘Press to Inhale Again’ screen

The ‘Press to Inhale again’ screen is displayed when an incorrect inhalation has been detected, and no dose has been delivered. You should inhale again by pressing the ON button.

 

Attention! to improve the quality of the inhalation, exhale before inhaling, then place your lips on the mouthpiece and inhale at a uniform rate for 3 seconds until the inhalation indicator light at the back of the inhaler turns off.

If an inhalation was not performed within 30 seconds of preparing a dose, a ‘No inhalation occurred’ screen will appear, and the inhalation indicator light will turn off. To return to the ‘Ready for inhalation’ screen, press the ON button and continue following the instructions. If within 7 seconds, the ON button has not been pressed – the inhaler will turn itself off.

* In the event that inhalation was not performed, the dose will be kept for future use.